Model ME2 Packaging

Instrument Applications:
- Nondestructive Air Mass Extraction Closure Integrity testing for:
- Pharmaceutical blister packs, pouches, vials and more.
- Beverage packaging (bottles, cans, carts etc.)
- Food - flexible and rigid packages
- Fully automatic Leak Testing under vacuum.
- Accommodates very large defect (to detect open package)
- Large defect and fine defect test features.
- Leak Tightness from 1x10-5 sccs at 2 psia (1 µm defect size); or
- Leak Tightness from 5x10-7 sccs at 0.02 psia (0.1 µm defect size)
- cleanroom or industrial applications
Leak Test Simulator: Use our Leak Test Simulator to optimize your application and define your micro-flow sensor/instrument.
Available Options: Please see ATC's options chart.
Features:

- All ME2 basic features
- Automated vacuum test circuit, uniquely designed for ultra-tight leak spec.
- Proprietary design for very large leak detection, for head-space of 0.4 cc and larger.
- Stainless steel chamber, optimized for package type, protects package from over expansion.
- Ergonomic design, manual or fully automated parts loading/unloading.
- Rugged design, very simple to operate.
- Includes SS Vacuum chambers and test cart.
- Pass/Fail, with automated test data acquisition.
- Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS).
Applicable Micro-Flow Sensors (IGLS):
- Vacuum Sensor: Model IL2-M, calibrated from 2 psia (slip and viscous flow regimes)
- Hard Vacuum Sensor: Model IMFS, calibrated from 0.01 to 0.2 psia (molecular and transitional flow regimes)
- Very Large leak detectors (available only for this product)
Specifications
Dimensions & Access:
- 17”W, 24”H, 19”D (Bench top size does not include connectors, fittings or chamber)
- Optional Cart Size: 28”W, 47”, 20”D
- Vacuum Chamber: Stainless steel with volume fillers / flexible part support
- Expansion tank, vacuum generation and control package are externally mounted on optional cart
Gases:
- Dry, clean gases: air, nitrogen
- Non-corrosive to stainless steel
- For other gases, consult ATC, Inc.
- Digital Inputs: 5 VDC, NPN Opto-Isolated for start, stop, type, pressure
switch, verify. Digital Outputs: 30 VDC-20 mA, NPN, opto-isolated for
pass, fail, clamp, test type, exhaust and custom
Analog outputs: single channel, 0-5 VDC pressure control - Pneumatic connections:
Test Ports
Air supply: ¼” Swagelok
Vacuum: 3/8” to 1” NW
(Application dependent) - RS232 Serial Interface
Print Model ME2 Packaging - PDF (594K)
Top of Page