Instrument Applications:
- Medium and low flow range measurement
- Portable transfer standard
- Calibration of Leak Test orifices and instruments
- Measurement of gas mixtures
- Pressure up to 500 PSIG
Features:
- Micro-Flow Sensor - The Intelligent Gas Flow Sensor (IGFS)
- Measures volume flow - using ATC's patented ALF design
- Dual range calibration, automatic range switches
- Measures absolute downstream pressure and temperature
- Optional: Compensation for various gas properties, temperature and viscosity
- Stainless steel rugged construction
- No moving parts
- No over-flow sensitivity
- Max. Differential Pressure: Up to 500 psi
Flow Ranges:

- 0-1 mL/min
- 0-3 mL/min
- 0-5 mL/min
- 0-10 mL/min
- 0-25 mL/min
- 0-50 mL/min
- 0-100 mL/min
- 0-250 mL/min
- 0-500 mL/min
Sensor Type: Accelerated Laminar Flow (ALF).
Measurement Uncertainty: ±1% (0.5% optional) of reading, volumetric flow, calibrated.
Pressure Range:
- 0-15; 30; 100; 250 (psi-absolute)
- Type: Absolute, micro-machined
- Measurement Uncertainty: 0.2% of Full Scale (FS)
- Maximum over Pressure: 1.2 times full range/scale
Temperature Range:

- Operating and Calibrated: 10 to 45°C (Higher temperature available, consult ATC, Inc.)
- Storage: -25 to 50°C
- Sensor Type: RTD 100 Ohms
- Measurement uncertainty at calibrated range: 0.5°C
Resolution: 16 bits A/D and 16 bits D/A
Interface:
- Serial port
- Digital I/O: Start/Stop, type clamp and more…
- Analog I/O
- Power: 115 or 220 VAC, single phase
- Optional: 2-line character display
Max. Differential Pressure: 500 psid (DP) or max. range of pressure sensor, whichever is smaller.
Gases Used:
- Use dry non-condensing and clean gases
- Air, Nitrogen, Helium, Argon, & Carbon Dioxide, Hydrogen, Natural Gas, Moisturized Gases
- Other Gases Available, consult ATC, Inc.
Response Time:
- Step function: from no flow to full range - 50 ms or less
- Sensor only, no volume
Print Model IF2-C / IF2-L - PDF (383K)
Top of Page