Model E2

Instrument Applications:
- Leak testing under pressure (max. 160 psig)
- Leak rates of 0.1 cc/min and higher
- Medium sized parts, higher throughput
- Cleanroom or industrial applications
- User friendly operator interface - stand alone applications
Leak Test Simulator: Use our Leak Test Simulator to optimize your application and define your micro-flow sensor/instrument.
Available Options: Please see ATC's options chart.
Features:
- Micro-Flow Sensor - The Intelligent Gas Leak Sensor (IGLS)
- Automated pressure test circuit with Balance and Quick-Fill features
- Front TFT graphical color display with touch screen
- Built in Equivalent Diameter (ED)
- Digital and analog I/O Interface
- Ethernet or serial interface
- Stainless steel enclosure for cleanroom
- Multiple test profiles
Applicable Micro-Flow Sensors (IGLS):
- Model IL2-M at pressure to 10 psig
- Model IL2-C
Specifications
Dimensions:
- 12”W, 12”H, 12”D (Does not include connectors & fittings)
- Expansion tank and pressure regulators are externally mounted
Gases:
- Dry, clean gases: air, nitrogen
- For other gases, consult ATC, Inc.
- Power Supply:
115VAC-60Hz
220VAC-50Hz (optional) - Pneumatic connections:
Test Ports
Air supply: ¼” Swagelok - Digital Inputs: 5 VDC, NPN Opto-Isolated for start, stop, type, pressure switch, verify
Digital Outputs: 30 VDC-20 mA, NPN,
Opto-isolated for pass, fail, clamp, test type, exhaust and custom
Analog outputs: single channel, 0-5 VDC pressure control - User selected
RJ45 Ethernet or
RS232 Serial Interface
Print Model E2 - PDF (762K)