Model IL2-HP
Home > Micro-Flow Sensors > Sensor IL2-HP

Sensor IL2-HP

The IL2-HP sensor is designed for use with the following products: IPE, IPE-HP.

Product Usage

  • High Pressure applications
  • Range: 500 – 2200 psig
  • Leak Rates: 0.05-2 cc/min

Flow Range:

  • 0-1 cc/min (mL/min)
  • 0-3 cc/min (mL/min)
  • 0-5 cc/min (mL/min)
  • 0-10 cc/min (mL/min)
  • 0-25 cc/min (mL/min)

Interface:

  • Serial port
  • Digital I/O: Start/Stop, type clamp and more…
  • Analog I/O
  • Power: 115 or 220 VAC, single phase
  • Optional: 2-line character display

Pressure Range:

  • 0-500; 1000; 2000; (psi-absolute)
  • Type: Absolute, micro-machined
  • Maximum over Pressure: 1.2 times full range/scale

Temperature Range:

  • Operating and Calibrated: 10 to 45°C
  • Storage: -25 to 50°C
  • Sensor Type: RTD 100 Ohms

Resolution:

16 bits A/D and 16 bits D/A

Features

  • Micro-Flow Sensor – The Intelligent Gas Leak Sensor (IGLS)
  • Measures volume flow – using ATC’s patented ALF design
  • Stainless steel rugged construction
  • No moving parts
  • No over-flow sensitivity
  • Max. Differential Pressure: Up to 500 psi DP
  • Optional dual range calibration

Sensor Type:

Accelerated Laminar Flow (ALF).

Measurement Uncertainty:

  • Flow: ±2% of reading, calibrated range.
  • Pressure: 0.3% of Full Scale (FS)
  • Temperature uncertainty at calibrated range: 0.5°C

Max. Differential Pressure:

500 psid (DP) or max. range of pressure sensor, whichever is smaller.

Gases Used:

  • Use dry non-condensing and clean gases
  • Air, Nitrogen, Helium, Argon, & Carbon Dioxide
  • Other Gases Available, consult ATC

Response Time:

  • Step function: from no flow to full range – 50 ms or less
  • (Open outlet, dependent on sensor range)
All products ship fully calibrated. No daily calibration required.